高级检索

    热脱附-气相色谱-质谱法测定胶粘剂中苯系物的释放量

    Determination of Benzene Series Emission in Adhesives by GC-MS with Thermal Desorption

    • 摘要: 将胶粘剂样品置于模拟实际环境的气候箱中,在280 ℃条件下解吸5 min,脱附的苯系物气体引入DB-1MS毛细管色谱柱进行分离并进入质谱仪进行测定。质谱分析中采用电子轰击离子源(EI)和选择离子监测(SIM)模式。6种苯系物的质量在10.73~1 216 ng之间呈线性,最低检出浓度(以采样体积5.0 L计)为0.2 μg·m-3。用标准加入法测得6种苯系物的回收率在87.8%~105%之间,测定值的相对标准偏差(n=8)在8.2%~11%之间。

       

      Abstract: The adhesive sample was placed into an environmental chamber where the actual environment was simulated. In the chamber, the sample was desorbed at 280 ℃ for 5 min. The benzene homologs in the sample were liberated and the gaseous mixture was introduced into the DB-1MS capillary chromatographic column for separation, after which the analytes were determined by MS. EI and SIM were adopted in MS. Linearity ranges were found between 10.73 to 1 216 ng, for 6 homologs, with minimum detection concentration of 0.2 μg·m-3 (based on 5.0 L of sampling volume). Recovery was tested by standard addition method, giving results ranged from 87.8%-105% for the 6 homologs. Values of RSD′s (n=8) found were in the range of 8.2%-11%.

       

    /

    返回文章