    离子色谱法测定水利工程水样中Cl-和SO42-的含量

    IC Determination of Cl- and SO42- in Water Samples of Hydraulic Engineering

    • 摘要: 取水利工程水样过滤后,以TSKgel IC-Anion-PWXL PEEK阴离子柱为分离柱,以2.5 mmol·L-1四硼酸钠溶液(用乙二胺调节其酸度至pH 10.5)为淋洗液进行离子色谱分离,用非抑制型电导检测器进行测定。Cl-和SO42-的质量浓度均在2.50~50.00 mg·L-1内与其对应的峰面积呈线性关系,Cl-和SO42-的检出限(3S/N)依次为5.1,7.6 μg·L-1。对空白水样连续测定5次,Cl-和SO42-的峰面积的相对标准偏差依次为4.1%,2.3%。测得加标回收率在97.9%~102%之间。

       

      Abstract: Water sample of hydraulic engineering was filtered, and separated by IC, using TSKgel IC-Anion-PWXL PEEK anionic column as separation column, and 2.5 mmol·L-1 sodium tetraborate solution (pre-adjusted to pH 10.5 with ethylenediamine) as eluant. Non-restrain conductance detector was used in the determination. Linear relationships between values of peak area and mass concentration of Cl- and SO42- were kept in the same range of 2.50-50.00 mg·L-1, with detection limits (3S/N) of 5.1, 7.6 μg·L-1 respectively. Precision of the method was tested with blank water sample for 5 determinations, values of RSD's (in terms of peak area) found for Cl- and SO42- were 4.1% and 2.3%, respectively. Values of recovery found by standard addition method were in the range of 97.9%-102%.

       

    /