高级检索

    气相色谱-质谱法测定生活饮用水中23种有机磷农药的含量

    GC-MS Determination of 23 OPPs in Drinking Water

    • 摘要: 将搅拌棒置于20 mL顶空瓶中,向顶空瓶中加入水样18 mL和甲醇2 mL,盖紧盖子,以转速800 r·min-1于20℃下搅拌溶液20 min。取出搅拌棒,将其放入热脱附玻璃管中。解吸液以DB-5MS色谱柱为固定相进行气相色谱分离。质谱分析中采用选择离子监测模式。23种有机磷农药的质量浓度在0.05~5.00 mg·L-1范围内与其峰面积呈线性关系,检出限(3S/N)为0.3~25 μg·L-1。按标准加入法在3个浓度水平上进行回收试验,回收率在70.9%~97.1%之间,测定值的相对标准偏差(n=6)均小于5.5%。

       

      Abstract: The stir bar was placed in 20 mL of a headspace bottle, and 18 mL of water sample and 2 mL of methanol were added into the headspace bottle. After closed the lip, the solution was stirred for 20 min at 20℃ with the speed of 800 r·min-1. The stir bar was taken out and placed into the thermal desorption glass tube. Desorption solution was separated by GC using DB-5MS chromatographic column as stationary phase. SIM was adopted in MS analysis. Linear relationships between values of peak area and mass concentration of 23 OPPs were found in the range of 0.05-5.00 mg·L-1, with detection limits (3S/N) in the range of 0.3-25 μg·L-1. Test for recovery was made by standard addition at 3 concentration levels, giving results in the range of 70.9%-97.1%, with RSDs (n=6) all less than 5.5%.

       

    /

    返回文章
    返回