    连续流动分析法测定电子烟烟液中的烟碱

    Determination of Nicotine in E-Cigarette Liquid by Continuous Flowing Analysis

    • 摘要: 称取电子烟烟液样品0.100 0 g于25 mL具盖离心管中,加入15%(体积分数)异丙醇溶液10 mL,溶解混合,静置后作为样品溶液。如果在静置过程中出现不完全混溶现象,可将离心管置于涡旋振荡器内,以5 000转·min-1的转速振荡1 min,离心后,取澄清液体作为样品溶液。用连续流动分析仪测定样品溶液在波长460 nm处的吸光度。烟碱的质量浓度在10.0~500 mg·L-1内与其对应的吸光度呈线性关系,测定下限(10s)为1.5 mg·L-1。方法用于测定市售电子烟烟液中的烟碱,测定值与标记值基本相符,加标回收率在99.5%~101%之间,测定值的相对标准偏差(n=6)为0.66%。

       

      Abstract: The e-cigarette liquid sample (0.100 0 g) was added into a 25 mL centrifuge tube with cover, and 10 mL of 15% (volume fraction) isopropanol solution was added to disolve the sample. After mixing well, the mixture was let stand to obtained the sample solution. If incomplete miscibility occured during standing, the centrifuge tube was oscillated at 5 000 r·min-1 for 1 min in a vortex oscillator. After centrifugation, the clear liquid was taken as the sample solution. The absorbance of the sample solution at the wavelength of 460 nm was measured using a continuous flowing analyzer. Linear relationship was found between absorbance and mass concentration of nicotine in the range of 10.0-500 mg·L-1, with lower limits of determination (10s) of 1.5 mg·L-1. The proposed method was applied to determination of nicotine in liquid of commercially available e-cigarette, giving measured values basically consistent with the measured values. Values of recovery obtained by standard addition method were in the range of 99.5%-101%, and RSDs (n=6) was 0.66%.

       

    /