高级检索

    应用磁性氧化石墨烯作为吸附分离净化剂-气相色谱-质谱法测定植物油中16种邻苯二甲酸酯类塑化剂含量

    GC-MS Determination of 16 Phthalates in Plant Oil with Magnetic Graphene Oxide as Adsorbent in Separation and Purification

    • 摘要: 气相色谱-质谱法(GC-MS)测定植物油中16种邻苯二甲酸酯类塑化剂(PAEs)的第一步是提取样品中的PAEs。取样品0.500 0 g,加入正己烷100 μL和乙腈2 mL,超声提取5 min,离心5 min,收集上清液,于留存的下层液相中再加乙腈2 mL,重复提取1次。所得上清液与第一次上清液合并在40℃氮吹蒸发至近干。于残渣中加入乙腈1 mL使溶解。在此溶液中依次加入无水硫酸镁250 mg,N-丙基乙二胺(PSA)80 mg和磁性氧化石墨烯50 mg,超声萃取5 min,通过外加磁场收集经净化的上清液,供GC-MS分析。在GC分析中,采用RXI-5si1MS色谱柱,按程序升温(在80~280℃区间)进行色谱分离,进样量为1 μL。在MS分析中采用电子轰击离子源(EI)和选择离子监测(SIM)模式。用基质匹配法绘制标准曲线,测得16种PAEs的线性范围均在0.02~1.00 mg·L-1之间,其检出限(3S/N)为0.005~0.008 mg·kg-1。以空白植物油为基体,用标准加入法进行回收试验,测得回收率在82.2%~111%之间。16种PAEs测定值的相对标准偏差(n=6)在1.0%~7.2%之间。应用此方法分析了市购的5种植物油样品,并在样品中检出5种PAEs。

       

      Abstract: The 1st step to the determination of 16 phthalates (PAEs) in plant oil by GC-MS was to separate the analytes from oil sample. 0.500 0 g of the sample was taken and extracted ultrasonically with 100 μL of n-hexane and 2 mL of acetonitrile for 5 min. The mixture was centrifuged for 5 min and the supernatant was collected. The lower phase remained was re-extracted with 2 mL of acetonitrile, and the 2nd supernatant obtained was combined with the 1st one. The combined supernatants were evaporated by N2-blowing at 40℃ to near dryness, and the residue was dissolved with 1 mL of acetonitrile. To this solution, 250 mg of anhydrous MgSO4, 80 mg of PSA and 50 mg of magnetic graphene oxide were added in succession, and the mixture was extracted ultrasonically for 5 min. Phase separation was attained and the supernatant was collected by external magnetic field and was to be used as sample solution for GC-MS analysis. An aliquot of 1 μL of the sample solution was introduced into the gas chromatograph and the 16 PAEs in it were separated on the chromatographic column of RXI-5si1MS under the condition of programed temperature elevation in the interval from 80℃ to 280℃. EI ionization source and mode of SIM were adopted in MS analysis. Matrix matching was used in preparation of standard curves, and linearity ranges found for the 16 PAEs were same between 0.02-1.00 mg·L-1, with their detection limits (3S/N) ranged from 0.005-0.008 mg·kg-1. Using blank plant oil as matrixes, test for recovery was made by standard addition method, giving values of recovery in the range of 82.2%-111%. RSDs (n=6) found for the 16 PAEs were in the range of 1.0%-7.2%. Five samples of plant oil were analyzed by the present method, and 5 PAEs were found in separate samples.

       

    /

    返回文章
    返回