Abstract:
Recycle delay time was set as 50 s on the basis of
13C longitudinal relaxation time(1.4-24.6 s) of potassium glycine-CO
2 absorption system obtained with pulse sequence t1irpg.
13C-qNMR method was established for determination of CO
2 loading of potassium glycine-CO
2 absorption system with inverse gated decoupling technique (pulse sequence Zgig30). The method was applied to the analysis of several samples, and compared with the traditional titration method. The method has the advantages of high accuracy, environmental friendliness, and the possibility of identifying species change according to chemical shift.