高级检索

    电感耦合等离子体质谱法测定化妆品中汞元素的含量

    Determination of Mercury Element in Cosmetics by Inductively Coupled Plasma Mass Spectrometry

    • 摘要: 针对电感耦合等离子体质谱法(ICP-MS)测定汞元素时存在汞记忆效应严重,且消除时前处理过程复杂的问题,提出了此项研究。取样品0.500 0 g置于聚四氟乙烯消解罐中,加入5 mL硝酸和1 mL 30%(质量分数)过氧化氢溶液,置于100 ℃恒温消解仪中加热30 min,冷却,将消解罐密封置于微波消解仪中,消解后冷却,打开消解罐,置于100 ℃恒温消解仪中15 min,将消解溶液转移至25 mL容量瓶中,用水洗涤消解罐数次,清洗液与消解溶液合并,用水定容至25 mL,与清洗试剂2 mg·L−1金元素标准使用液或2%(质量分数,下同)L-半胱氨酸溶液配制的20 μg·L−1186Re内标溶液在线混匀后,采用ICP-MS进行测定。结果表明,清洗试剂可消除0.5~5 μg·L−1汞元素标准溶液的汞记忆效应,对下一个样品的测定结果不产生影响。汞元素的质量浓度在0.5~5 μg·L−1内与所对应的信号强度与内标元素信号强度比呈线性关系,采用2 mg·L−1金元素标准使用液配制内标溶液时,检出限(3s)为0.001 0 mg·kg−1,采用2% L-半胱氨酸溶液配制内标溶液时,检出限(3s)为0.001 2 mg·kg−1。按照标准加入法进行回收试验,回收率为90.2%~104%,测定值的相对标准偏差(n=6)均小于9.0%。方法用于3种市售化妆品分析,使用2种清洗试剂配制的内标溶液得到的检出量一致,为0.008 0~0.13 mg·kg−1

       

      Abstract: In view of the serious mercury memory effect and the complicated elimination pretreatment process in determination of mercury element by inductively coupled plasma mass spectrometry (ICP-MS), this study was proposed. 0.500 0 g of sample was taken and placed in a polytetrafluoroethylene digestion tank, and 5 mL of nitric acid and 1 mL of 30% (mass fraction) hydrogen peroxide solution were added. The mixture was heated at 100 ℃ for 30 min in a constant temperature digestion instrument and cooled. The digestion tank was sealed and placed in a microwave digestion instrument. After digestion, the digestion tank was cooled and opened, and placed at 100 ℃ for 15 min in a constant temperature digestion instrument. The digestion solution was taken and placed in a 25 mL-volumetric flask, and the digestion tank was washed several times with water. The washing solutions were combined with the digestion solution, and the solution was made up to 25 mL with water. The 20 μg·L−1 186Re internal standard solution, which was prepared with cleaning reagents 2 mg·L−1 gold element standard solution or 2% (mass fraction, the same below) L-cysteine solution, was mixed online with the sample solution for determination by ICP-MS. As shown by the results, the cleaning reagents could eliminate the mercury memory effect of 0.5−5 μg·L−1 mercury element standard solution, without affecting the determination of the next sample. Linear relationships between the ratios of the corresponding signal intensity to the internal standard signal intensity and mass concentrations of mercury element were found in the range of 0.5−5 μg·L−1, with detection limit (3s) of 0.001 0 mg·kg−1 by using 2 mg·L−1 gold element standard solution to prepare an internal standard solution, and detection limit (3s) of 0.001 2 mg·kg−1 by using 2% L-cysteine solution to prepare an internal standard solution. Test for recovery was made by the standard addition method, giving results in the range of 90.2%−104%, with RSDs (n=6) of the determined values less than 9.0%. This method was used for the analysis of 3 commercially available cosmetics, and the detection amounts, which obtained from the internal standard solution prepared by two cleaning reagents, were consistent and in the range of 0.008 0−0.13 mg·kg−1.

       

    /

    返回文章
    返回