高级检索

    顺序注射荧光光谱法测定环境水中痕量西维因

    Sequential Injection Fluorescence Spectroscopy Determination of Carbaryl in Environmental Water

    • 摘要: 利用农药西维因与β-环糊精在pH 5.0的磷酸氢二钠-柠檬酸缓冲溶液中反应生成主客体包合络合物,导致西维因本身的特征荧光增强的现象,结合顺序注射进样方式提出了顺序注射荧光光谱法测定环境水样中痕量西维因的方法。试验得出仪器的主要工作条件:① 进样方式:两区带;② 环糊精的浓度:3.5 mmol·L-1;③ 试样体积:240 μL;④ 缓冲溶液体积:100 μL;⑤ 测定流量:5.0 mL·min-1;⑥ 扫描速率:12 000 nm·min-1;⑦ 激发波长:280 nm,发射波长:334 nm。在设定条件下,西维因的质量浓度在0.90~210 μg·L-1范围内与相对荧光强度呈线性关系。检出限(3s)为0.3 μg·L-1。用此方法测定了河水、湖水和自来水样品中西维因含量,并在此3种样品的基体中加入标准溶液进行回收试验,测得回收率在91.8%~109%之间。

       

      Abstract: Based on the fact that β-cyclodextrin forms a subject-object complex in Na2HPO4-citric acid buffer solution of pH 5.0 with carbaryl by embracing carbaryl in its inner cavity resulting enhancement of the inherent fluorescence of carbaryl, and by incorporating sequential injection technique of sample introduction, a method of sequential injection fluorescence spectroscopy determination of carbaryl in environmental water samples was proposed. Optimum working conditions for the instrument were given as follows: ① mode of sample introduction: di-band; ② concn. of β-CD in the reaction mixture: 3.5 mmol·L-1; ③ vol. of sample solution: 240 μL; ④ vol. of buffer solution: 100 μL; ⑤ flow-rate of sample solution: 5.0 mL·min-1; ⑥ scanning rate: 12 000 nm·min-1; and ⑦ wavelengths of excitation: 280 nm, and of emission: 334 nm. Under the precribed conditions, linear relationship between values of relative fluorescence intensity and mass concentration of carbaryl was obtained in the range of 0.90-210 μg·L-1. Detection limits (3s) was found to be 0.3 μg·L-1. Samples of river, lake and tap water were analyzed by this method, and test for recovery was performed by addition of standard solution of carbaryl to these samples as matrixes, giving results of recovery in the range from 91.8% to 109%.

       

    /

    返回文章
    返回