Advanced Search
    SHI Ting-ting, SI Xiong-yuan, TAN Hua-rong, SI You-bin, LIU Wei. Application of Biphenyl-Molecularly Imprinted Polymers in GC Determination of Biphenyl and 3,3-Dichlorobiphenyl[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2016, 52(8): 869-873. DOI: 10.11973/lhjy-hx201608001
    Citation: SHI Ting-ting, SI Xiong-yuan, TAN Hua-rong, SI You-bin, LIU Wei. Application of Biphenyl-Molecularly Imprinted Polymers in GC Determination of Biphenyl and 3,3-Dichlorobiphenyl[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2016, 52(8): 869-873. DOI: 10.11973/lhjy-hx201608001

    Application of Biphenyl-Molecularly Imprinted Polymers in GC Determination of Biphenyl and 3,3-Dichlorobiphenyl

    • Waste water sample was treated by self-made biphenyl-molecularly imprinted polymers, and GC was applied to the determination of biphenyl and 3,3-dichlorobiphenyl in the sample solution. The analytes were separated on ZB-624 capillary column (30 m×0.25 mm, 0.25 μm) and detected by FID. Linear relationships between values of peak area and mass concentration of biphenyl and 3,3-dichlorobiphenyl were kept in the same range of 0.20-20 mg·L-1, with detection limits (3S/N) of 20.0, 35.0 μg·L-1 respectively. Recovery of the method was tested by standard addition method, and values of recovery and RSD′s (n=5) found were in the ranges of 91.2%-98.5% and 4.7%-5.9% respectively.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return