Advanced Search
    FAN Bolun, MA Zhenzhen, TIAN Huan, BAO Zhengyu, ZHOU Yang, ZHANG Hongyu. HG-AFS Determination of Selenium in Plants with Closed Digestion[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2017, 53(8): 905-907. DOI: 10.11973/lhjy-hx201708008
    Citation: FAN Bolun, MA Zhenzhen, TIAN Huan, BAO Zhengyu, ZHOU Yang, ZHANG Hongyu. HG-AFS Determination of Selenium in Plants with Closed Digestion[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2017, 53(8): 905-907. DOI: 10.11973/lhjy-hx201708008

    HG-AFS Determination of Selenium in Plants with Closed Digestion

    • HG-AFS was applied to the determination of selenium in plants with closed digestion. The plant sample was digested closedly for 4 h with the mixture of HNO3-H2O2 (2+1) at 130℃. 10 g·L-1 NaBH4 solution was used as reductant. Linear relationship between values of fluorescence intensity and mass concentration of selenium was kept in the range of 2-10 μg·L-1, with detection limit (3s/k) of 0.18 μg·L-1. Precision of the method was tested at the concentration level of 10 μg·L-1 selenium standard solution for 6 determination and value of RSD found was 0.63%. On the base of standard material of cabbage (GBW 10014), test for recovery was made by standard addition method. Value of recovery found was 102%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return