Simultaneous ICP-MS Determination of 17 Elements in Urine with Dynamic Reaction Cell
-
Graphical Abstract
-
Abstract
The urine sample was acidized with HNO3 to 1% of volume fraction of HNO3. After centrifugation, the supernatant was diluted 10 times with 4%(φ) HNO3 solution. ICP-MS was applied to the simultaneous determination of 17 elements, i.e. Pb, Cd, As, Se, Mn, Cu, Zn, Sb, Be, V, Fe, Ni, Mo, Cr, Tl, Co and Sn in the solution with dynamic reaction cell. 45Sc (for determination of Mn、Cu、Be、V、Fe、Ni、Cr、Co), 72Ge (for determination of As、Se、Zn), 115In (for determination of Cd、Sb、Mo、Sn) and 209Bi (for determination of Pb、Tl) were used as internal standards. Linearity range of Fe found was in the range within 5.00 mg·L-1and linearity ranges of the others were in the range within 50.0 μg·L-1, with detection limits (3s) in the range of 0.000 5-0.496 0 μg·L-1. The proposed method was applied to the analysis of urine sample, giving values of recovery and RSD's (n=7) in the ranges of 85.6%-110% and 0.68%-4.8% respectively.
-
-