GC-MS Determination of N-Vinyl-2-Pyrrolidone in Plastic, Electronic and Electrical Products
-
Graphical Abstract
-
Abstract
A method was proposed for the determination of N-vinyl-2-pyrrolidone (NVP) in plastic, electronic and electrical products by gas chromatography-mass spectrometry (GC-MS). The sample of 1.000 0 g was extracted ultrasonically with tetrahydrofuran of 10 mL for 30 min, and the extract was filtered through PTFE membrane and then determined by GC-MS using DB-624 capillary column for temperature programming. The quanlitative ions (m/z 56, 111, 28, 41) and the quantitative ion (m/z 111) were selected in SCAN and SIM, respectively. linear relationship between values of peak area and mass concentration of NVP was obtained in range of 0.05-10 mg·L-1 with detection limit(3S/N) of 0.1 mg·kg-1. Test for recovery was made by standard addition method, giving results in the range of 94.9%-114%, with RSDs (n=7) ranged from 0.040% to 1.6%.
-
-