    • Spectrophotometry was applied to the determination of rhenium in roasted sample of molybdenum concentrate. The optimized conditions found were as follows:① wavelength of determination was 426 nm; ② coloration in HCl (1+1) solution; ③ amount of 350 g·L-1 potassium rhodanate solution was 1.0 mL; ④ amount of 20 g·L-1 stannous chloride solution was 1.0 mL; ⑤ amount of H2SO4 was 3 mL; ⑥ heating temperature was 85℃; ⑦ heating time was 1.5 h; ⑧ amount of rhenium reagent was 1.0 mL; ⑨ pH of the sample solution was about 9. Linear relationship between values of absorbance and mass concentration of rhenium was obtained in the range of 3.0-30.0 mg·L-1, with detection limit (3s/k) of 1.0 mg·L-1. The proposed method was applied to the analysis of roasted sample of molybdenum concentrate, giving results in consistency with the values obtained by ICP-AES with RSD's (n=7) in the range of 1.1%-2.2%. Test for recovery was performed by standard addition method, giving values of recovery in the range of 94.0%-106%.