    KONG Weisong, YANG Yekun, LU Sheming, XU Yuqiong, GENG Yongqin, MIAO Enming, YANG Guangyu, LI Jing. Determination of Benzo[a]pyrene and 3-Hydroxybenzo[a]pyrene in Urine of Smokers by Ultra High Performance Liquid Chromatography Combined with a Novel Pretreatment Technique[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2018, 54(6): 621-626. DOI: 10.11973/lhjy-hx201806001
    Citation: KONG Weisong, YANG Yekun, LU Sheming, XU Yuqiong, GENG Yongqin, MIAO Enming, YANG Guangyu, LI Jing. Determination of Benzo[a]pyrene and 3-Hydroxybenzo[a]pyrene in Urine of Smokers by Ultra High Performance Liquid Chromatography Combined with a Novel Pretreatment Technique[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2018, 54(6): 621-626. DOI: 10.11973/lhjy-hx201806001

    Determination of Benzoapyrene and 3-Hydroxybenzoapyrene in Urine of Smokers by Ultra High Performance Liquid Chromatography Combined with a Novel Pretreatment Technique

    • A novel sample pretreatment device integrating extraction, purification and concentration was designed and fabricated on the base of Soxhlet extraction device, and by practice of matrix solid-phase dispersion extraction and column chromatography, and it was used for sample pretreatment in the determination of benzoapyrene (BaP) and 3-hydroxybenzoapyrene (3-OHBaP) in urine of smokers by ultra high performance liquid chromatography (UPLC). 50 mL of urine sample were collected and hydrolyzed vortically with 0.5 g of acetate buffer solution and 0.1 mL of β-glucuronidearyl sulfatase at 38℃ for 5 h. An aliquot (20.00 mL) of the hydrolyzed solution was concentrated to near dryness and dispersed with 2.0 g of silica gel. The dispersion, after drying, was introduced into the novel sample pretreatment device as prescribed, extracted and purified by refluxing at 90℃ in a water bath for 45 min using 60 mL of a mixture of cyclohexane and ethanol (98+2) as extractant. Heating was continued until the solution was evaporated to near-dryness. The residue was taken up with 1 mL of acetonitrile and the solution, after filtration, was used for UPLC analysis. ACQUITY UPLCTM BEH C18 column was used as stationary phase and a mixture of acetonitrile and water (68+32) was used as mobile phase to elute the analytes. Fluorophotometric detections were made for 3-OHBaP in the interval of peak-appearing time from 0 to 6.5 min at λex 365 nm and λem 450 nm, and for BaP in the time interval from 6.5 to 12.0 min at λex 384 nm and λem 406 nm. Linear relationships between values of peak area and respective mass concentration were found in the range of 4.0 to 200 μg·L-1 for 3-OHBaP and in the range of 10 to 250 μg·L-1 for BaP, with detection limits (3S/N) of 0.65 μg·L-1 and 3.6 μg·L-1 respectively. Values of recovery obtained by standard addition method were in the ranges of 86.8% to 92.8% and 91.2% to 95.7% respectively. Values of RSDs (n=7) were all less than 4.0%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return