Advanced Search
    YANG Jing, CAO Yang, DONG Jiao, WANG Yueming. Determination of Tungsten in Alloys by X-Ray Fluorescence Spectrometry[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(6): 683-686. DOI: 10.11973/lhjy-hx201906012
    Citation: YANG Jing, CAO Yang, DONG Jiao, WANG Yueming. Determination of Tungsten in Alloys by X-Ray Fluorescence Spectrometry[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(6): 683-686. DOI: 10.11973/lhjy-hx201906012

    Determination of Tungsten in Alloys by X-Ray Fluorescence Spectrometry

    • XRFS was applied to the determination of tungsten in various tungsten containing alloy stees, Ni-based and Co-based alloys. The spectral line of Lβ1 of W was chosen as analytical spectral line, and 2θ of 39.50° was selected as background site. Spectral line of Lβ2 of Ta was found to interfere with the line of Lβ1 of W. Two sets of standard samples were selected and fundamental parameter (FP) method was used with help of FP software to find out the overlap-interfering coefficient k, which was listed in FP software to eliminate the overlap-interference from Ta to W.5 standard samples with W content ranging from 0.054% to 11.71%, were analyzed by the proposed method, giving results in consistency with the certified values, and RSDs (n=7) ranged from 0.55% to 9.5%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return