    Determination of H2O2 Based on Cytochrome C Double-Channel Ratiometric H2O2 Biosensor

    • Cytochrome C and mercaptopropionic acid modified Au electrode (Cyt. C/MPA/Au) and hexa-(ferrocene)-hexyl mercaptan modified Au electrode (FcHT/Au) were prepared, and the electrochemical property of two modified electrodes was studied by cyclic voltammetry. It was found that in PBS of pH 7.0, the reduction peak current of H2O2 at the Cyt. C/MPA/Au increased with the increase of the content of H2O2, while the reference peak current at the FcHT/Au remained basically unchanged. Based on this finding, a double-channel ratiometric H2O2 biosensor was proposed. Content of H2O2 was determined by DPV. Linear relationship between values of peak current ratio of H2O2 at Cyt. C/MPA/Au to that at FcHT/Au and concentration was ranged from 20.0-100.0 μmol·L-1, with detection limit (3S/N) of 1.19 μmol·L-1. Recovery of the method was in the range of 71.4%-79.2%.