    ZHOU Baolong. GC-MS/MS Determination of 12 UV Absorbers in Sunscreening Cosmetics[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(9): 1013-1018. DOI: 10.11973/lhjy-hx201909005
    Citation: ZHOU Baolong. GC-MS/MS Determination of 12 UV Absorbers in Sunscreening Cosmetics[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(9): 1013-1018. DOI: 10.11973/lhjy-hx201909005

    GC-MS/MS Determination of 12 UV Absorbers in Sunscreening Cosmetics

    • A portion (5.0 g) of the sunscreening cosmetic sample was mixed thoroughly with 1 mL of saturated NaCl solution and extracted twice with CH2Cl2 (5 mL for each extraction) to transfer 12 UV absorbers (UVA) into the organic solvent by swirling for 60 s, extracting ultrasonically for 30 min and centrifuging for 5 min with high speed. The supernatants from the 2 extractions were combined and evaporated to near dryness by N2-blowing. The residue was dissolved and made its total volume to 1.0 mL with CH2Cl2. An aliquot of 1 μL of the CH2Cl2 solution was introduced to the chromatographic column of DB-5MS for separation of the 12 UVA under the mode of programed temperature elevation in the temperature interval from 80℃ to 300℃. A clear-cut separation of the 12 UVA was fulfilled within 30 min and their determination was carried out by MS/MS under the modes of EI-ionization source and selected reaction monitoring. Linearity ranges for the 12 UVAs were found same between 0.01-1.0 mg·L-1. Values of their detection limits (3S/N) found were in the range of 1.0 to 3.0 μg·kg-1. Test for recovery and precision was performed by addition of mixed standard solution of the 12 UVA to a blank sunscreening cosmetic sample, giving results of recovery in the range from 93.0% to 100%, and results of RSDs (n=6) in the range from 1.2% to 4.1%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return