Advanced Search
    LIU Wei, LÜ Ting, TAO Meijuan. ICP-MS Determination of 19 Trace Elements in Graphene Powder Materials with Microwave Digestion[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(11): 1318-1322. DOI: 10.11973/lhjy-hx201911013
    Citation: LIU Wei, LÜ Ting, TAO Meijuan. ICP-MS Determination of 19 Trace Elements in Graphene Powder Materials with Microwave Digestion[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2019, 55(11): 1318-1322. DOI: 10.11973/lhjy-hx201911013

    ICP-MS Determination of 19 Trace Elements in Graphene Powder Materials with Microwave Digestion

    • The sample of graphene powder material was dissolved with the mixture of nitric acid, hydrofluoric acid and hydrogen peroxide by microwave digestion. After cooling, the digestion solution was diluted to 100.0 mL. ICP-MS was applied to the determination of 19 trace elements, i.e. As, Ba, Bi, Co, Cr, Cu, K, Mg, Mo, Ni, Pb, Sb, Sn, Ti, V, Zn, Cd, La and Ce in the sample solution. 103Rh and 45Sc were used as internal standards and isotopes were used to overcome mass spectrometry interference. Linear relationships between values of emission intensity and mass concentration of the 19 trace elements were kept in the same range within 500 μg·L-1, with detection limits in the range of 0.01-3.9 μg·L-1. Values of recovery found by standard addition method were in the range of 82.2%-123%, and values of RSDs (n=8) of determined values were in the range of 2.6%-8.5%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return