    HE Ying, LI Jinsong, TONG Wenfeng, XU Qiao, WANG Tianqi, YAN Qing, LI Jian, ZHOU Yanxiao. Determination of Residues of 19 Perfluoroalkyl Compounds in Water Quality by Online Dual Solid Phase Extraction-Ultra-High Performance Liquid Chromatography-Triple Quadrupole Tandem Mass Spectrometry[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2025, 61(8): 951-959. DOI: 10.11973/lhjy-hx240379
    Citation: HE Ying, LI Jinsong, TONG Wenfeng, XU Qiao, WANG Tianqi, YAN Qing, LI Jian, ZHOU Yanxiao. Determination of Residues of 19 Perfluoroalkyl Compounds in Water Quality by Online Dual Solid Phase Extraction-Ultra-High Performance Liquid Chromatography-Triple Quadrupole Tandem Mass Spectrometry[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2025, 61(8): 951-959. DOI: 10.11973/lhjy-hx240379

    Determination of Residues of 19 Perfluoroalkyl Compounds in Water Quality by Online Dual Solid Phase Extraction-Ultra-High Performance Liquid Chromatography-Triple Quadrupole Tandem Mass Spectrometry

    • The water sample was passed through a 0.45 µm glass fiber filter membrane, and an aliquot (10 mL) was taken. Then, the 100 µL of 1 µg·L-1 mixed internal standard solution was added. After mixing well, the mixed solution was passed through a 0.22 µm disposable polyethersulfone syringe filter. Perfluoroalkyl compounds (PFCs) in the filtrate were enriched using online dual solid phase extraction system and then introduced into ultra-high performance liquid chromatograph-triple quadrupole tandem mass spectrometer. Separation was performed on Waters BEH C18 column using the gradient elution system consisting of acetonitrile and 2 mmol·L-1 ammonium acetate solution. Mass spectrometric analysis was conducted in negative ion mode of electrospary ion (ESI) source with multiple reaction monitoring (MRM) mode, and quantification was achieved using the internal standard method. It was showed that linear relationships of the mass concentrations of the 19 PFCs and ratios of their peak areas to internal standard peak areas were kept in definite ranges, with detection limits (3S/N) in the range of 0.01-0.03 ng·L-1. Spike recovery tests were performed using surface water, groundwater, drinking water, and industrial wastewater as matrices, giving recoveries in the range of 84.1%-118%, and RSDs (n=6) of the determined values ranged from 2.2% to 6.8%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return