IC Determination of Trace Amounts of Bromate and Nitrite in Bottled or Barrelled Purified Water
-
Graphical Abstract
-
Abstract
Contents of bromate and nitrite in bottle or barrelled purified water were determined by ion chromatography (IC). The protective column (AG19-4mm) and chromatographic column (DIONEX AS19) were used for separation, and KOH solutions of different concentrations were used as eluent. The bromate and nitrite were separated successfully. Linearity ranges of BrO3- and NO2- were found between 5.00-100.00 μg·L-1 and 1.00-50.00 μg·L-1, with detection limits (3s/k) of 0.08 and 0.14 μg·L-1 respectively. The proposed method was applied to the determination of BrO3- and NO2- in the samples of bottled or barrelled purified water, values of recovery found by standard addition method were in the range of 92.7%-98.4%, and RSD′s (n=5) were less than 5.0%.
-
-