Stripping Voltammetric Determination of Silver Ion with Modifed Carbon Fiber Electrode as Working Electrode
-
Graphical Abstract
-
Abstract
Method of preparation and pretreatment of carbon fiber electrode (CFE) was described,and the CFE obtained was modified by electrochemical polymerization of the monomer thiophene.In the determination of Ag+-ion,the test solution in 0.1 mol·L-1 HNO3 was pretreated by reduction for 120 s at -0.5 V,where the Ag+-ion was complexed with polythiophene.Silver ion in the complex was then stripped from the electrode by differential pulse anodic stripping voltammetry with scanning in the range from 0 to 0.6 V,and the oxidation peak potential of silver ion was observed at 0.24 V,where linear relationship between values of peak current of Ag+-ion and its concentration was obtained in the range of 5.0×10-7-1.0×10-3mol·L-1.Detection limit (3S/N) of the method found was 5.0×10-8mol·L-1.
-
-