    FANG Han-chen, LI Juan. FI-HG-Electrothermo-atomic Absorption Spectrometric Determination of Traces of Mercury in water[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2007, 43(12): 1068-1069.
    Citation: FANG Han-chen, LI Juan. FI-HG-Electrothermo-atomic Absorption Spectrometric Determination of Traces of Mercury in water[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2007, 43(12): 1068-1069.

    FI-HG-Electrothermo-atomic Absorption Spectrometric Determination of Traces of Mercury in water

    • In an acidic solution,mercury of different states present in water sample was oxidized to divalent state by potassium permanganate or bromate.The method,by hyphenation of flow-injection technique and hydride generation with thermo-atomic absorption spectrometry,was used for the determination of mercury in the sample solution.Linearity range of the method was found in the range of 0.2-15 μg·L-1 with correlation coefficient of 0.999 9,when measured at the wavelength of 253.7 nm.Detection limit (3σ) of the method was found to be 0.094 μg·L-1.Test for recovery was made by standard addition method on the base of substantial water sample,giving values of recovery in the range of 95.7% to 104.6%.