Advanced Search
    HU Hong-mei, GUO Yuan-ming, LI Tie-jun, YOU Ju-ju, SUN Xiu-mei, ZHONG Zhi, HE Yi-na. GC Determination of 11 Chlorobenzenes in Aquaculture Water with Static Headspace[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2015, 51(1): 25-30.
    Citation: HU Hong-mei, GUO Yuan-ming, LI Tie-jun, YOU Ju-ju, SUN Xiu-mei, ZHONG Zhi, HE Yi-na. GC Determination of 11 Chlorobenzenes in Aquaculture Water with Static Headspace[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2015, 51(1): 25-30.

    GC Determination of 11 Chlorobenzenes in Aquaculture Water with Static Headspace

    • GC was applied to the determination of 11 chlorobenzenes in aquaculture water with static headspace. The optimum experimental conditions found were to put 10 mL water sample containing 200 g·L-1 sodium chloride in a 20 mL vial and stir at 70 ℃ for 30 mins. 11 chlorobenzenes were separated completely by DB-35 MS capillary column (30 m×0.25 mm, 0.25 μm), detected by electron capture detector and then quantify through external standard method or standard addition method. The linearity of the method ranged from 0.16 to 8.0 μg·L-1 for dichlorobenzene isomers, 0.017 6-0.88 μg·L-1 for trichlorobenzene isomers, 0.004-0.2 μg·L-1 for tetrachlorobenzene isomers and from 0.001 to 0.05 μg·L-1 for pentachlorobenzene and hexachlorobenzene. The limits of detection ranged from 0.0002 to 0.04 μg·L-1. The proposed method was used in the analysis of aquaculture water samples, giving values of recovery and RSD′s (n=5)in the ranges of 86.0%-106% and 2.1%-5.8% respectively.