Capillary Column-GC Determination of Residual Amount of Ethylene Glycol in Ethylene Glycol Stearate with Fusion Extraction
-
Graphical Abstract
-
Abstract
Residual amount of ethylene glycol in ethylene glycol stearate was determined by capillary column GC. The sample was melted at 50 ℃ ultrasonically and extracted with methanol. Propylene glycol was used as internal standard. An aliquot of the extract was separated on HP-5 capillary column and determined with FID. Linearity range of ethylene glycol was kept from 102.6 to 1 026.0 mg·L-1, with value of detection limit (3S/N) of 30 mg·L-1. The proposed method was applied to the analysis of ethylene glycol stearate, values of recovery found were in the range of 93.7%-103.2%.
-
-