Advanced Search
    SHEN Song-li, LU Heng-jun, WEI Chao, CHEN Mei-lan. IC Determination of Phenol in Environmental Water with Hollow Fibrous-Three Phase Liquid Phase Micro-extraction[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2013, 49(1): 63-65.
    Citation: SHEN Song-li, LU Heng-jun, WEI Chao, CHEN Mei-lan. IC Determination of Phenol in Environmental Water with Hollow Fibrous-Three Phase Liquid Phase Micro-extraction[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2013, 49(1): 63-65.

    IC Determination of Phenol in Environmental Water with Hollow Fibrous-Three Phase Liquid Phase Micro-extraction

    • IC was used in the determination of phenol in environmental water. The sample was extracted by hollow fibrous membranes-three phase liquid phase micro-extraction with n-octyl alcohol as extraction solvent. The extraction was carried out under magnetic stirring for 40 min at the rate of 600 r·min-1. The SH-2 Anion column was used for separation, and 12 mmol·L-1 NaOH solutions was used as eluent. Linearity range of phenol was found between 1.00-500 μg·L-1, with detection limits (3S/N) of 0.8μg·L-1. The proposed method was applied to the determination of phenol in surface water, values of recovery found by standard addition method were in the range of 91.0%-109%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return