Advanced Search
    LI Fen, LI Qi-hua, FAN Chao-ying, MA Ying. ICP-AES Determination of Scandium in Mg-Sc Alloy[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2011, 47(7): 816-817.
    Citation: LI Fen, LI Qi-hua, FAN Chao-ying, MA Ying. ICP-AES Determination of Scandium in Mg-Sc Alloy[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2011, 47(7): 816-817.

    ICP-AES Determination of Scandium in Mg-Sc Alloy

    • An ICP-AES method for the determination of Sc in Mg-Sc alloy was proposed. The sample was dissolved with HCl (1+1) solution and H2O2. The wavelength of 361.384 nm was selected as analytical spectral line for the determination of Sc. Linear relationship between values of emission intensity and mass concentration of Sc was obtained in the range within 20.00 mg·L-1, with detection limit (3s) of 2.9×10-4mg·L-1. Recovery and precision of the method were tested by analyzing 4 Mg-Sc alloy samples by standard addition method, giving values of recovery in the range of 98.0%-102.0% and RSD′s (n=6) in the range of 0.7%-1.8%.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return