ICP-AES for Determination of Trace of Hafnium in Nuclear-Pure Spongy Zirconium
-
Graphical Abstract
-
Abstract
ICP-AES method for the determination of trace Hf in nuclear-pure spongy Zr with standard addition calibration was experimentally studied.Determination range of Hf was 50-400 μg·g-1 in case concentration of Zr in the sample solution was 10 g·L-1.Results of recovery were in range of 103%-110% and values of RSD′s were in the range 4.2%-6.5%.
-
-