GC-MS Determination of Volatile Organic Compounds in Ground Water with Purger and Trap
-
Graphical Abstract
-
Abstract
The PTA 3000 Purge and Trap was used in combination with GC-MS determination of 54 volatile organic compounds (VOC′s) in ground water.Temperature and time for purge and trap were set at 40 ℃ and 11 min respectively.Ten mL.of sample solution were transferred to the purge and trap bottle and introduced into the instrument for GC-MS analysis.The VOC′s were separated on the DB-5MS GC-column and detected by the full-scanning mode with positive electron ionization.Selective ion mass was used in the quantification by internal standard.Detection limit (3S/N) found for all the VOC′s tested were in the range of 0.03-0.12 μg·L-1.Tests for recovery were made by addition of standard VOC′s at 3 different concentration levels to sample matrixes,values of recovery obtained were in the range of 83.2%-115.5%.Values of RSD′s (n=7) found for the VOC′s were in the range of 0.85%-3.88%.
-
-