刘子琪,等:液相色谱-高分辨质谱法定性定量检测血液透析器中3种未知可沥滤物


                                   t
                   and  tri(2-ethylhexyl)rimellitate  from  haemodialysis   北京:中国标准出版社,2008.
                   tubing[J]. International Journal of Pharmaceutics,2001,   [14]  国家食品药品监督管理总局. 血液透析及相关治疗血
                   229(1/2):139-146.                                 液透析器、血液透析滤过器、血液滤过器和血液浓缩器:
               [12]  刘子琪,陈卓颖,刘璟,等. 气相色谱-三重四极杆                        YY 0053—2016[S]. 北京:中国标准出版社,2018.
                   质谱法测定血液透析器中的可沥滤物[J]. 分析试验                     [15]  国家市场监督管理总局,国家标准化管理委员会. 人体
                   室,1-7[2022-08-30].  DOI:  10. 13595/j.cnki.issn1000-  血液及血液成分袋式塑料容器  第1部分: 传统型血袋:
                   0720. 2020. 033105.                               GB 14232. 1—2020[S]. 北京:中国标准出版社,2020.
               [13]  中华人民共和国国家质量监督检验检疫总局,中国国家                    [16]  陈华燕,徐苏华,黄敏菊. 高效液相色谱荧光法测定
                   标准化管理委员会. 医用输液、输血、注射器具检验方                         血液透析器中双酚A溶出量[J]. 化学分析计量,2022,
                   法  第1部分: 化学分析方法:GB/T 14233. 1—2008[S].            31(5):8-13.

                Qualitative and Quantitative Detection of 3 Unknown Leachable Substances in

               Hemodialyzers by Liquid Chromatography-High Resolution Mass Spectrometry


                                            LIU Ziqi , HUANG Yuanli , FU Bufang 1*
                                                    1,2
                                                                    1
              (1. National Institutes for Food and Drug Control, Beijing 102629, China; 2. Shandong Lab of Advanced Biomaterials and Medical
                                               Devices in Weihai, Weihai 264209, China)

                  Abstract: Referring to the hemodialyzer instructions and GB/T 14232. 1—2020, water and ethanol-water mixed solution
              were used to extract leachable substances from the hemodialyzer. The chromatography and mass spectrometry information of unknown
              leachable substances in the extract was collected by liquid chromatography-high resolution mass spectrometry. The qualitative process
              was introduced using dimethylacetamide as an example, and the content of leachable substances was determined by external standard
              method. It was shown that 3 leachable substances were found around 1. 7, 2. 1, 2. 5 min, which were determined to be 2-pyrrolidone,
              dimethylacetamide, and N-methylpyrrolidone after mass spectrometry analysis and comparison with reference substances. Linear
              relationships between values of the mass concentration of 3 leachable substances and the corresponding peak area of quantitative ions
              were kept in the ranges of 0. 089 6−4. 48 mg · L  −1 , 0. 469 5−18. 78 mg · L  −1 , and 0. 102 6−5. 130 mg · L  −1 , with detection
                                                   −1
              limits (3S/N) of 0. 005 1, 0. 009 9, 0. 010 5 μg · g .  The method was used for the analysis of 28 hemodialyzers, and the leaching
              amounts of 3 leachable substances in ethanol-water extract were higher than those in water extract, with the maximum leaching
              amounts of 376. 7, 2 668. 4, 517. 8 μg, respectively.
                  Keywords: hemodialyzer; liquid chromatography-high resolution mass spectrometry; leachable substance